Date : 10/31/2022 6:28:22 PM
From : "Arnon Rubinshtein"
To : "Ronald Berger-Lefébure" , "eyaly"
Cc : "Eyal Yahel"
Subject : Re: Follow up LIBS Issues
Attachment : 8445_image001.jpg;

Hello Ronald,

Please see files in following link:
https://drive.google.com/drive/folders/1aLdegcYrkZcYEUlTKMy0bZwA2sghCrSR?usp=share_link

Feel free to connect me for more data or explanations.

Best regards,

Arnon A. Rubinshtein
NRCN, P.O. Box 9001
Beer-Sheva 84190, Israel.
Email: arnonr@nrcn.gov.il
Tel: +972-50-6244294.

From: Ronald Berger-Lefébure <ronald.bergerlefebure@iumtek.com>
Sent: Monday, October 31, 2022 3:37:05 PM
To: eyaly <eyaly@nrcn.gov.il>
Cc: Arnon Rubinshtein <arnonr@nrcn.gov.il>; Eyal Yahel <eyalyahel@gmail.com>
Subject: Follow up LIBS Issues
 

Dear Eyal and Arnon,

 

Thank you for your message and I am fine, so far so good.

 

Regarding your request, the best is to schedule a Teams/Skype after receiving data from your side, in order to be sure to understand your issues. We will also come back to you with our thoughts and support.

 

With my best regards,

 

Ronald

 

 

De : eyaly <eyaly@nrcn.gov.il>
Envoyé : samedi 29 octobre 2022 21:31
À : Ronald Berger-Lefébure <ronald.bergerlefebure@iumtek.com>
Cc : Arnon Rubinshtein <arnonr@nrcn.gov.il>; Eyal Yahel <eyalyahel@gmail.com>
Objet : LIBS Issues

 

Dear Ronald, 

 

I hope this message finds you all well.

We would to share a problem we encountered and appreciate your advice on the performance of the instrument we purchased.

 

As you probably recall, the instrument has been received in Israel, in January 2021 but was installed only in April 2022 by iumtek personnel (Covid-19 issues). 

During the installation, instrument operation was demonstrated with a single Aluminum standard.

 

Since then, we have learned how to operate the instrument for its basic features. 

Next, we tested multiple standards (Aluminum mostly) to build and calibrate an analytical method for impurities measurements and acquire a database for Aluminum matrixes.

 

During the development process, we noticed the following:

1.       Instrument sensitivity is poor and prevents measuring low-level impurities.

2.       Repeatability of the signal (at consecutive laser shots) is not good as expected.

 

In addition, the auto-focusing function doesn’t work very well.

Clearly, we need to resolve these issues and appreciate suggestions how to improve sensitivity and repeatability. 

 

I suggest that Arnon (cc'd) will send you some typical spectrums of Aluminum standards (maybe one that has been supplied with the instrument and additional ones). We can also send certification of analysis.

 

Best regards. 

Eyal

 

Eyal Yahel, Ph.D.
Chief Scientist, NRCN
P.O.Box 9001 Beer-Sheva, 
Israel, 84190